ZM4 Instruction Manual 9R87-B June 2012
25
Volts/mA Auto Step Setup
Figure 20: Volts/mA Auto Step Setup Display
Reference Function
Start Select to start the Auto Step simulation.
Mode: Volts/mA mode for the simulation. For
reference only, cannot be modified here.
Begin Value: Starting simulation value. Brings up edit
display to change the value.
End Value: Ending simulation value. Brings up edit
display to change the value.
Num Steps: Number of steps in the procedure. Brings up
edit display to adjust the value between 1
and 10.
Dwell Time: Time to pause at each step before
advancing to the next step. Brings up edit
menu to select the value between Manual, 5
or 10 seconds.
Startup Delay:
(not shown in Fig 20)
Delay after pressing Start key before
simulation begins. Brings up edit menu to
select the value between None, 30 or 60
seconds.