ZM4 Instruction Manual 9R87-B June 2012
22
Volts/mA Sensor Setup
Figure 17: Volts/mA Sensor Setup Display
Reference Function
Mode Choose Volts Measure, mA Measure, 24V
Source, Volts Source, mA Source, or mA
Sink
Damping Enable or disable damping and adjust value.
Active setting shown to the right.
Calibration Perform a Zero, Field Calibration, or Restore
Factory Defaults
Reset Min/Max Resets Min/Max value
Resolution Choose Accuracy or Precision resolution on
display for selected sensor. Active setting
shown to the right.
Information Displays Sensor Type, Accuracy, Calibration
Date, Serial # and Range
NOTE:
Description of Mode setup is described in the following pages. Damping and
Calibration are described in the preceding pages covering Pressure Sensor
setup.